Standard operating procedure reduces interoperator variation and improves accuracy when measuring packed cell volume

Abstract

Objective To evaluate whether a standard operating procedure (SOP) for canine packed cell volume (PCV) measurement reduces operator-dependent variation and improves accuracy within a veterinary teaching hospital environment.

Materials and methods Clinical staff and final-year undergraduate veterinary students were recruited to perform PCV measurements in blinded duplicate samples. Participants were randomly allocated to perform this with or without an SOP. Participants’ results were compared against a reference, generated by the authors following the World Health Organization guidelines.

Results The study population comprised 18 clinical staff and 39 students. Three clinical staff and seven students displayed errors consistent with inaccurate reading, only one of whom had access to the SOP. Five students and two clinical staff had errors attributable to incorrect preparation, with only one having access to the SOP. Interoperator variation was significantly less using the SOP. Using the SOP, 95 per cent of the results were within 0.0125 l/l of the reference value, in comparison with within 0.09 l/l without SOP. Interoperator variation was significantly less in the SOP group (P=0.0025).

Clinical significance Using the SOP resulted in less variation and more accurate results. This confirms that PCV measurement with an SOP can truly be a ‘waived’ test.

  • PCV
  • packed cell volume
  • standard operating procedure
  • SOP
  • variation
View Full Text

Statistics from Altmetric.com

Request Permissions

If you wish to reuse any or all of this article please use the link below which will take you to the Copyright Clearance Center’s RightsLink service. You will be able to get a quick price and instant permission to reuse the content in many different ways.