A spectral-domain optical coherence tomography device provides reliable corneal pachymetry measurements in canine eyes
  1. A. F. Alario, DVM and
  2. C. G. Pirie, DVM, DACVO
  1. Department of Clinical Sciences, Tufts Cummings School of Veterinary Medicine, North Grafton, MA 01536, USA
  1. E-mail for correspondence: Chris.pirie{at}tufts.edu

Abstract

The objective of this study was to determine central corneal thickness (CCT) and the intra- and interuser reliability using a portable spectral-domain optical coherence tomography (SD-OCT) device in canine eyes. Twenty clinically normal dogs were examined. CCT measurements were obtained from both eyes of each animal three times by two operators in succession. The CCT was automatically calculated using the pachymetry software generated from eight radial scans, 6 mm in length. Mean canine CCT was 606.83±39.45 μm for all eyes examined. There was no significant difference in CCT based on the eye examined (OD vs OS), age or gender of the animal. There was no significant difference in CCT between replicates performed by the same operator; however, a small but significant difference was noted in CCT between operators. The mean difference in CCT between operators was 1.9 μm (P=0.03). The coefficient of variation for each user and between users was very low (range 0.64–1.7 per cent). The intraclass correlation coefficient comparing operators was 0.975. Based on these results, the SD-OCT device evaluated is capable of obtaining precise CCT measurements with excellent intra- and interoperator reliability in canine eyes.

  • Accepted April 29, 2013.

Statistics from Altmetric.com

  • Accepted April 29, 2013.
View Full Text

Request permissions

If you wish to reuse any or all of this article please use the link below which will take you to the Copyright Clearance Center’s RightsLink service. You will be able to get a quick price and instant permission to reuse the content in many different ways.