Consistency of results from the Vettest 8008 clinical biochemistry analyser
  1. CJ Little,
  2. G Gettinby,
  3. J Downs and
  4. CB Marshall

Abstract

Findings are presented on the precision of a clinical biochemistry analyser designed specifically for veterinary use. Twenty biochemical analytes have been examined in detail for variation within and between runs. The results indicate that the analyser can provide high precision for all the analytes with the possible exception of calcium, and suggest that the instrument can be used with confidence in the practice laboratory to aid diagnosis and to monitor biochemical changes in animals receiving treatment.